Introduction
This page provides work instructions for on-site integration between the Kanomax Laser Particle Sensor and Element-D.
Materials
Your package contains the following items:
- Element-D w/ Power Supply
- Special wire
Setup Instructions
1. Plug in the Element-D and complete the Wifi Connect instructions.
2. Download the Kanomax LPS Setting Software onto a computer. Connect it to your Kanomax Laser Particle Sensor through the cable given and configure the settings to match exactly the photo below.
Press the Send button and wait for a confirmation message to pop up before disconnecting.
3. Connect the USB end of the cable to the Element-D and the RS-485 end to the Kanomax interface, circled below.
4. Turn on the Kanomax by plugging in the power source and attaching it to the vacuum source. Ensure that both the Particle and Status lights are on or else the Element-D will not be able to capture data.
5. Check your Elemental Machines dashboard account to confirm that particle count values are displayed.
Warning: The Element-D only captures data for particle count use. The capture of concentration data is not supported. Unplug the Element-D from the pH meter when capturing concentration data.