Introduction
This page provides work instructions on how to setup the Element-C to connect to the NuAire NU-5830, NU-5831, NU-5731, NU-8631 incubator models.
Materials
Your package contains the following items:
- Element-C w/Power Supply
- Connection cable
NOTE: Both of these components are unique to the NuAire NU-5830, NU-5831, NU-5731, NU-8631 incubator models.
Equipment Needed for Setup
- Small flathead screwdriver
Setup Instructions
- Carefully remove the contents from its packaging. If any of the contents looks damaged, please send a picture of the contents to help@elementalmachines.com.
- Power on the Element-C and complete the Wifi Connect instructions.
- Connect the open wires from the Element-C cable to the green terminal block on the Element-C:
- Slightly loosen the screws for Slots 1-6 on the Element-C terminal block.
- Insert the BROWN Wire from Connection Cable into Slot 1. Tighten the screw to ensure that the wire is secure.
- Insert the BROWN-WHITE Wire from Connection Cable into Slot 2. Tighten the screw to ensure that the wire is secure.
- Insert the ORANGE Wire from Connection Cable into Slot 3. Tighten the screw to ensure that the wire is secure.
- Insert the ORANGE-WHITE Wire from Connection Cable into Slot 4. Tighten the screw to ensure that the wire is secure.
- Insert the GREEN Wire from Connection Cable into Slot 5. Tighten the screw to ensure that the wire is secure.
- Insert the GREEN-WHITE Wire from Connection Cable into Slot 5. Tighten the screw to ensure that the wire is secure.
Figure 1: Cable wire connection into Element-C
4. Connect the RJ-45 cable from the Element-C to the NuAire incubator 4-20mA data port. This can be found on the back of the incubator on the top left corner, as shown in Figure 2 below.
Figure 2: Insert the RJ-45 jack into the 4-20mA port on the NuAire incubator
5. Confirm that the incubator conditions appear on the Elemental Machines dashboard. Set up any desired alerts.