Introduction
This page provides work instructions for on-site integration between the Kuhner Shaker NET60 Modbus RTU model and Element-D.
Materials
Your package contains the following items:
- Element-D w/Power Supply
- RS-232 to Male USB to serial cable
- Mounting plate
Prerequisites
Prior to following these setup instructions:
1. The Kuhner Net-60 Modbus RTU, specifically configured for either the X or Z series shaker, must be installed on each shaker.
2. You must have an adapter for the cable. This can purchased from Kuhner, PN EED060421.
Equipment Needed for Setup
- Small Screwdriver
Setup Instructions
1. Carefully remove the contents from its packaging. If any of the contents looks damaged, please send a picture of the contents to help@elementalmachines.com.
2. Plug in the Element-D and complete the Wifi Connect instructions. Unplug the Element-D when complete.
3. Unscrew the 4 screws on the side of the NET60 enclosure and lift off the enclosure.
4. Set the Serial connection baud rate, stop bit and parity to 19.2 kilobits/second, 2 and none and the node ID to 1 by setting the 12 DIL switches on the NET60 module
to the following values
where 1 is the DIL switch is in the up position (away from the enclosure)
5. Put the enclosure on the NET60 and screw in the 4 screws on the sides of the enclosure.
6. Add the adapter to the RS-232 end of the cable. Connect the USB end of the cable to the Element-D and the RS-232 end to the to the NET60 module.
7. Power on the Element-D.
8. Check your Elemental Machines dashboard account to confirm that incubator shaker values are displayed.
9. Setup desired alerts for the shaker on the Elemental Machines dashboard.