Introduction

This page provides work instructions for on-site integration between the Kuhner Shaker NET60 Modbus RTU model and Element-D.


Materials

Your package contains the following items:

  • Element-D w/Power Supply
  • RS-232 to Male USB to serial cable
  • Mounting plate


Prerequisites

Prior to following these setup instructions:

1. The Kuhner Net-60 Modbus RTU, specifically configured for either the X or Z series shaker, must be installed on each shaker. 

2. You must have an adapter for the cable.  This can purchased from Kuhner, PN EED060421.


Equipment Needed for Setup

  • Small Screwdriver


Setup Instructions

1. Carefully remove the contents from its packaging.  If any of the contents looks damaged, please send a picture of the contents to help@elementalmachines.com.

2. Plug in the Element-D and complete the Wifi Connect instructions. Unplug the Element-D when complete. 

3. Unscrew the 4 screws on the side of the NET60 enclosure and lift off the enclosure.

4. Set the Serial connection baud rate, stop bit and parity to 19.2 kilobits/second, 2 and none and the node ID to 1 by setting the 12 DIL switches on the NET60 module

to the following values


where 1 is the DIL switch is in the up position (away from the enclosure)


5. Put the enclosure on the NET60 and screw in the 4 screws on the sides of the enclosure.

6. Add the adapter to the RS-232 end of the cable. Connect the USB end of the cable to the Element-D and the RS-232 end to the to the NET60 module.


7. Power on the Element-D.

8. Check your Elemental Machines dashboard account to confirm that incubator shaker values are displayed.

9. Setup desired alerts for the shaker on the Elemental Machines dashboard.