Applicable ThermoFisher Cryo Plus Liquid Nitrogen Storage System Series

  • CryoPlus 1/2/3/4, Series 7400


Package Contents

  • Element-B
  • Connection Cable


Safety Information

Review and abide by the safety information found in the Element-B and ThermoFisher Cryo Plus 7400 Series Liquid Nitrogen Storage System user manuals. Tidy connection cable to prevent tripping hazards.


Installation Instructions

Receiving the Shipment

1. If this is your first time setting up the Elemental Machines system, you will receive an email for your dashboard account verification. Save this email for when the devices arrive.

2. When devices are shipped, they will be added to your Elemental Insights Dashboard™ with default names. When you first log in you will see all your devices with a 'disconnected' status.

3. When devices arrive, carefully remove the contents from the packaging. If any of the contents look damaged, please send a picture to help@elementalmachines.com

4. If you have ordered multiple Element-B’s, ensure the Element-B being used is configured as Normally Closed (NC). This can be verified by checking the device label, the dashboard under ‘Machine Model’, or the packing list sent with your shipment. 

a. Element-B’s can either be configured as Normally Closed (NC) or Normally Open (NO) by our team here at Elemental Machines. Different assets require different configurations. If you have ordered Element-B’s for multiple assets, you may have a mix and need to be aware of attaching the correctly configured device to the correct asset.

5. Ensure there is an Elemental Machines Gateway set up in the vicinity of the Element-B.

a. Note: ‘Vicinity’ is within wireless connection range. This is usually up to 30 meters but can depend on the lab’s layout.

Attaching Connection Cable to ThermoFisher Cryo Plus 7400 Series Liquid Nitrogen Storage System

6. Locate the Remote Alarm Connector (RJ-11 telephone-style jack) on the rear panel of the system.

Shape, rectangle

Description automatically generated

Figure 1 Remote Alarm Connector circled

7. Plug the RJ-11 connector end (shown in Figure 2) of the connection cable into the RJ-11 port on the system.

How to Connect Patch Panel with the Switch? - Copperled Technology Co., Ltd.

Figure 2 RJ-11 end of Element-B connection cable

Attaching Connection Cable to Element

8. Per your discretion, write the name and/or location of the equipment being monitored on the Element-B. This can be particularly helpful if multiple devices are being installed. Use of a Sharpie is recommended.

9. Plug the thermocouple end (Figure 3) of the connection cable into the Element-B.

a. Note Ensure the (-) and (+) prongs of the thermocouple end are properly matched.

Figure 3 Thermocouple end of Element-B connection cable

10. Pull the battery tab on the Element-B.

Positioning the Element-B

11. Determine where the Element-B should be placed so that it is not blocked, which allows a better wireless connection, and so that the connection cable is tidy.

12. Securely position the Element-B in the chosen location. The Element-B has magnets which make it easily mountable to magnetic surfaces. 

a. Note: If applicable, an adhesive metal mounting plate available from EM may be used.

Programming Alerts

13. Program alerts on the ThermoFisher Cryo Plus 7400 Series Liquid Nitrogen Storage System. Refer to the ThermoFisher Cryo Plus 7400 Series Liquid Nitrogen Storage System user manual for which alarm conditions the system can detect and how to set up alerts.

14. Set up alert rules on the Elemental Insights Dashboard™ for the Element-B. Guidance on how to do so is available in the Support section of the dashboard.


NOTE: Alerts must be programmed on both the ThermoFisher Cryo Plus 7400 Series Liquid Nitrogen Storage System and the Elemental Insights Dashboard™ to receive threshold alerts through the Elemental Machines platform.