Introduction
This page provides work instructions on how to setup the Element-C to connect to the Steri-Cult incubator. These instructions apply to Thermo Steri-Cult 3307 and 3310 models.
Materials
Your package contains the following items:
- Element-C w/Power Supply
- Connection cable NOTE: This is unique to the Steri-Cult Incubator
Equipment Needed for Setup
- Small flathead screwdriver
- Additional screwdriver needed for Thermo Steri-Cult incubator connection
Setup Instructions
- Carefully remove the contents from its packaging. If any of the contents looks damaged, please send a picture of the contents to help@elementalmachines.com.
- Plug in the Element-C and complete the Wifi Connect instructions (see attachment).
- Wire the Connection cable to Element-C.
- Slightly loosen the screws in Slots 1 through 6 on the green terminal block on the Element-C. Figure 1: Steri-Cult cable wiring into Element-C Terminal Block
- Insert the RED Wire from the Connection Cable into SLOT 1 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Insert the WHITE Wire from Connection Cable into SLOT 2 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Insert the BLUE Wire from Connection Cable into SLOT 3 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Insert the GREEN Wire from Connection Cable into SLOT 4 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Insert the BLACK Wire from Connection Cable into SLOT 5 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Insert the BROWN Wire from Connection Cable into SLOT 6 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Wire the Connection cable to the Steri-Cult Incubator:
- Slightly loosen the screws for the TEMP, CO2, and RH channels on the terminal block of the Thermo Steri-Cult Incubator.Figure 2: Connection Cable is wired into the TEMP, CO2, and RH (+) and (-) Channels of the Thermo Incubator
- Insert the RED Wire from Connection Cable into the Temp (+) slot. Tighten the screw to ensure that the wire is secure.
- Insert the WHITE Wire from Connection Cable into the Temp (-) slot. Tighten the screw to ensure that the wire is secure.
- Insert the BLACK Wire from Connection Cable into the CO2 (+) slot. Tighten the screw to ensure that the wire is secure.
- Insert the BROWN Wire from Connection Cable into the CO2 (-) slot. Tighten the screw to ensure that the wire is secure.
- Insert the BLUE Wire from Connection Cable into the RH (+) slot. Tighten the screw to ensure that the wire is secure.
- Insert the GREEN Wire from Connection Cable into the RH (-) slot. Tighten the screw to ensure that the wire is secure.