Introduction

This page provides work instructions on how to setup the Element-C to connect to the Steri-Cult incubator. These instructions apply to Thermo Steri-Cult 3307 and 3310 models.

Materials

Your package contains the following items:

  • Element-C w/Power Supply
  • Connection cable  NOTE: This is unique to the Steri-Cult Incubator

Equipment Needed for Setup

  • Small flathead screwdriver
  • Additional screwdriver needed for Thermo Steri-Cult incubator connection

Setup Instructions

  1. Carefully remove the contents from its packaging.  If any of the contents looks damaged, please send a picture of the contents to help@elementalmachines.com. 
  2. Plug in the Element-C and complete the Wifi Connect instructions (see attachment).
  3. Wire the Connection cable to Element-C. 
    1. Slightly loosen the screws in Slots 1 through 6 on the green terminal block on the Element-C.                                    Figure 1: Steri-Cult cable wiring into Element-C Terminal Block
    2. Insert the RED Wire from the Connection Cable into SLOT 1 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
    3. Insert the WHITE Wire from Connection Cable into SLOT 2 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
    4. Insert the BLUE Wire from Connection Cable into SLOT 3 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
    5. Insert the GREEN Wire from Connection Cable into SLOT 4 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
    6. Insert the BLACK Wire from Connection Cable into SLOT 5 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
    7. Insert the BROWN Wire from Connection Cable into SLOT 6 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
  4. Wire the Connection cable to the Steri-Cult Incubator:
    1. Slightly loosen the screws for the TEMP, CO2, and RH channels on the terminal block of the Thermo Steri-Cult Incubator.Figure 2: Connection Cable is wired into the TEMP, CO2, and RH (+) and (-) Channels of the Thermo Incubator
    2. Insert the RED Wire from Connection Cable into the Temp (+) slot. Tighten the screw to ensure that the wire is secure.
    3. Insert the WHITE Wire from Connection Cable into the Temp (-) slot. Tighten the screw to ensure that the wire is secure.
    4. Insert the BLACK Wire from Connection Cable into the CO2 (+) slot. Tighten the screw to ensure that the wire is secure.
    5. Insert the BROWN Wire from Connection Cable into the CO2 (-) slot. Tighten the screw to ensure that the wire is secure.
    6. Insert the BLUE Wire from Connection Cable into the RH (+) slot. Tighten the screw to ensure that the wire is secure.
    7. Insert the GREEN Wire from Connection Cable into the RH (-) slot. Tighten the screw to ensure that the wire is secure.