Introduction
This page provides work instructions on how to setup the Element-C to connect to the Thermo SteriCycle Series incubator. These instructions apply to the following Thermo SteriCycle models:
- 370
- 371
- 380
- 381
Materials
Your package contains the following items:
- Element-C w/Power Supply
- Connection cable NOTE: This is unique to Thermo SteriCycle Series Incubators
Equipment Needed for Setup
- Small flathead screwdriver
Setup Instructions
- Carefully remove the contents from its packaging. If any of the contents looks damaged, please send a picture of the contents to help@elementalmachines.com.
- Plug in the Element-C and complete the Wifi Connect instructions (see attached).
- Unplug the Element-C.
- Wire the Connection cable to Element-C.
- Slightly loosen the screws in Slots 1 through 6 on the green terminal block on the Element-C. Figure 1: Connection cable wiring into Element-C Terminal Block
- Insert the RED Wire from the Connection Cable into SLOT 1 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Insert the WHITE Wire from Connection Cable into SLOT 2 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Insert the BLUE Wire from Connection Cable into SLOT 3 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Insert the GREEN Wire from Connection Cable into SLOT 4 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Insert the BLACK Wire from Connection Cable into SLOT 5 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Insert the BROWN Wire from Connection Cable into SLOT 6 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Power off the SteriCycle incubator.
- Remove the two screws on the top front of the incubator as shown in Figure 2. Figure 2: Screws are visible from top down view of incubator
- Route the Connection cable through the Wire thru-port (back of incubator) to to the Analog Board (front of the incubator), as shown in Figure 3 below. Figure 3: Routing for Connection cable through the Wire thru-port
- Slightly loosen the screws for the TEMP, CO2, and RH (+) and (-)channels on the terminal block of the Analog board on the Thermo SteriCycle Incubator.
- Insert the RED Wire from Connection Cable around the TEMP (+) slot. Tighten the screw to ensure that the wire is secure.
- Insert the WHITE Wire from Connection Cable around the TEMP (-) slot. Tighten the screw to ensure that the wire is secure.
- Insert the BLACK Wire from Connection Cable around the CO2 (+) slot. Tighten the screw to ensure that the wire is secure.
- Insert the BROWN Wire from Connection Cable around the CO2 (-) slot. Tighten the screw to ensure that the wire is secure.
- Insert the BLUE Wire from Connection Cable into the RH (+) slot. Tighten the screw to ensure that the wire is secure.
- Insert the GREEN Wire from Connection Cable into the RH (-) slot. Tighten the screw to ensure that the wire is secure. Figure 4: Connection Cable is wired into the TEMP, CO2, and RH (+) and (-) Channels of the Thermo Incubator
- Power on the incubator.
- Check the Elemental Machines dashboard to ensure that the incubator data appears online. Set any desired alerts on the platform.