This page provides work instructions on how to setup the Element-C to connect to the Thermo Forma 3950 and 3951 Reach-in Incubator.
Materials
Your package contains the following items:
- Element-C with Power Supply
- Connection cable
NOTE: Both of these components are unique to the Thermo Forma 3950 and 3951 models.
Equipment Needed for Setup
- Small flathead screwdriver
- Additional screwdriver needed for Thermo Forma Reach-In Incubator
Setup Instructions
- Carefully remove the contents from its packaging. If any of the contents looks damaged, please send a picture of the contents to help@elementalmachines.com.
- Plug in the Element-C and complete the Wifi Connect instructions (see attachment).
- Unplug the Element-C.
- Power off the Thermo Forma Reach-In Incubator.
- Remove the top of the Incubator. Route the connection cable through the Auxiliary Hole for Cables on the back of the incubator, as shown in Figure 1. Figure 1: Location of the Auxiliary Hole for Cables on the back of the Incubator
- Locate the Analog Output board, shown in Figures 2 and 2b, in the Incubator. Figures 2a and 2b: Electronics layout in the Forma Incubator (left) and Analog Output board close-up (right)
- Slightly loosen the screws for Temp and CO2 (+) and (-) channels, circled in Figure 2b above, on the terminal block of the Thermo Incubator.
- Insert the RED Wire from Connection Cable into Temp (+). Tighten the screw to ensure that the wire is secure.
- Insert the WHITE Wire from Connection Cable into Temp (-) slot. Tighten the screw to ensure that the wire is secure.
- Insert the GREEN Wire from Connection Cable around the CO2 (+) slot from top to bottom. Tighten the screw to ensure that the wire is secure.
- Insert the BLACK Wire from Connection Cable into the CO2 (-) slot from top to bottom. Tighten the screw to ensure that the wire is secure.
- Make sure the Connection Cable is wired to the Element-C as shown below:
- Slightly loosen the screws in Slots 1 through 4 on the green terminal block on the Element-C. Figure 3: Connection cable wiring into Element-C Terminal Block
- Insert the RED Wire from the Connection Cable into SLOT 1 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Insert the WHITE Wire from Connection Cable into SLOT 2 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Insert the GREEN Wire from Connection Cable into SLOT 3 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Insert the BLACK Wire from Connection Cable into SLOT 4 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Power on the Thermo Forma Reach-In Incubator.