Introduction
This page provides work instructions on how to setup the Element-C to connect to the Caron 7400-25-33 Series incubator. These instructions apply to the following Caron models:
- 7400-25
- 7400-33
- 7401-25
- 7401-33
- 7402-25
- 7402-33
- 7403-25
- 7403-33
Materials
Your package contains the following items:
- Element-C w/Power Supply
- 4- and 6-wire Connection cable (only one is required based on analog accessory)
Equipment Needed for Setup
- Small flathead screwdriver
- Additional screwdriver needed for Caron incubator connection
Setup Instructions
- Carefully remove the contents from its packaging. If any of the contents looks damaged, please send a picture of the contents to help@elementalmachines.com.
- Plug in the Element-C and complete the Wifi Connect instructions (see attachment).
Pre-requisites for Integrating with the Caron incubator:
- Determine whether the incubator is equipped with OUTP302 or OUTP303. OUTP302 will allow for monitoring of TEMP and one of the following: either RELATIVE HUMIDITY OR CO2. OUTP303 will allow for monitoring of TEMP, RELATIVE HUMIDITY, and CO2.
- If the incubator is equipped with OUTP302, use the 4-wire connection cable. If the incubator is equipped with OUTP303, use the 6-wire connection cable.
- Ensure that the temp range is set to -50C to 100C under the Analog Output Screen. NOTE: this is the default setting. If changed, Element-C will not output accurate data!
Caron Incubator Integration
OUTP302:
- Locate the Analog output connections on the back of the incubator. Determine which two parameters will be monitored: TEMP + RH or CO2 (pick one of the latter).
- Wire up the incubator for TEMP:
- Loosen the screw for TEMP I(+) Slot. Insert the RED wire from the cable into the slot. Tighten the screw to secure the wire.
- Loosen the screw for the COM(-) Slot under TEMP. Insert the WHITE wire from the cable into the slot. Tighten the screw to secure the wire.
- Select one of the following two parameter for monitoring (RH or CO2):
Incubator Monitoring for RH | Incubator Monitoring for CO2 |
|
|
Incubator Wiring - Temp + RH | Incubator Wiring - Temp + CO2 | ||
Slot | Wire Color | Slot | Wire Color |
TEMP I(+) | RED | TEMP I(+) | RED |
COM(-) under TEMP | WHITE | COM(-) under TEMP | WHITE |
RH I(+) | GREEN | CO2 I(+) | GREEN |
COM(-) under RH | BLACK | COM(-) under CO2 | BLACK |
Element-C Wiring:
- Wire up the Element-C for TEMP:
- Loosen the screw for Slot 1. Insert the free end of the RED wire from the cable into the slot. Tighten the screw to secure the wire.
- Loosen the screw for Slot 2. Insert the free end of the WHITE wire from the cable into the slot. Tighten the screw to secure the wire.
- Wire up the Element-C for one of the two options:
Element-C RH Setup | Element-C CO2 Setup |
|
|
Element-C Wiring: Temp + RH | Element-C Wiring: Temp + CO2 | ||
Slot | Wire Color | Slot | Wire Color |
1 | RED | 1 | RED |
2 | WHITE | 2 | WHITE |
3 | GREEN | 5 | GREEN |
4 | BLACK | 6 | BLACK |
OUTP303:
- Locate the Analog output connections on the back of the incubator.
- Wire up the incubator for TEMP:
- Loosen the screw for TEMP I(+) Slot. Insert the RED wire from the cable into the slot. Tighten the screw to secure the wire.
- Loosen the screw for the COM(-) Slot under TEMP. Insert the WHITE wire from the cable into the slot. Tighten the screw to secure the wire.
- Wire up the incubator for RH:
- Loosen the screw for RH I(+) Slot. Insert the BLUE wire from the cable into the slot. Tighten the screw to secure the wire.
- Loosen the screw for the COM(-) Slot under RH. Insert the GREEN wire from the cable into the slot. Tighten the screw to secure the wire.
- Wire up the incubator for CO2:
- Loosen the screw for CO2 I(+) Slot. Insert the BLACK wire from the cable into the slot. Tighten the screw to secure the wire.
- Loosen the screw for the COM(-) Slot under CO2. Insert the BROWN wire from the cable into the slot. Tighten the screw to secure the wire.
Incubator Wiring - Temp + RH + CO2 | |
Slot | Wire Color |
TEMP I(+) | RED |
COM(-) under TEMP | WHITE |
RH I(+) | BLUE |
COM(-) under RH | GREEN |
CO2 I(+) | BLACK |
COM(-) under CO2 | BROWN |
Element-C Wiring:
- Wire up the Element-C for TEMP:
- Loosen the screw for Slot 1. Insert the free end of the RED wire from the cable into the slot. Tighten the screw to secure the wire.
- Loosen the screw for Slot 2. Insert the free end of the WHITE wire from the cable into the slot. Tighten the screw to secure the wire.
- Wire up the Element-C for RH:
- Loosen the screw for Slot 3. Insert the free end of the BLUE wire from the cable into the slot. Tighten the screw to secure the wire.
- Loosen the screw for Slot 4. Insert the free end of the GREEN wire from the cable into the slot. Tighten the screw to secure the wire.
- Wire up the Element-C for CO2
- Loosen the screw for Slot 5. Insert the free end of the BLUE wire from the cable into the slot. Tighten the screw to secure the wire.
- Loosen the screw for Slot 6. Insert the free end of the GREEN wire from the cable into the slot. Tighten the screw to secure the wire.
Element-C Wiring: Temp + RH + CO2 | |
Slot | Wire Color |
1 | RED |
2 | WHITE |
3 | BLUE |
4 | GREEN |
5 | BLACK |
6 | BROWN |