Introduction
This page provides work instructions on how to setup the Element-C to connect to the Thermo Forma 4100 Series incubator at the customer site. These instructions apply to the following Thermo Forma models:
- 4110
- 4111
- 4120
- 4121
- 4130
- 4131
- 4140
- 4141
Materials
Your package contains the following items:
- Element-C w/Power Supply
- Thermo Incubator Custom Cable
- Connection cable NOTE: This is unique to Thermo Forma 4100 Series Incubators
Equipment Needed for Setup
- Small flathead screwdriver
- Additional screwdriver needed for Thermo Forma 4100 Series incubator connection
Setup Instructions
Prerequisites:
- Analog output board must be installed on the Incubator
Connecting Element-C on the Thermo Incubator:
- Carefully remove the contents from its packaging. If any of the contents looks damaged, please send a picture of the contents to help@elementalmachines.com
- Plug in the Element-C and complete the Wifi Connect instructions (see attachment).
- Power off the ElementC.
- Wire the Connection cable to Element-C.
- Slightly loosen the screws in Slots 1 through 8 on the green terminal block on the Element-C. Figure 1: Connection cable wiring into Element-C Terminal Block
- Insert the RED Wire from the Connection Cable into SLOT 1 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Insert the WHITE Wire from Connection Cable into SLOT 2 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Insert the ORANGE Wire from Connection Cable into SLOT 3 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Insert the BLUE Wire from Connection Cable into SLOT 4 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Insert the GREEN Wire from Connection Cable into SLOT 5 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Insert the YELLOW Wire from Connection Cable into SLOT 6 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Insert the BLACK Wire from Connection Cable into SLOT 7 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Insert the BROWN Wire from Connection Cable into SLOT 8 of the green terminal block on the Element-C. Tighten the screw to ensure that the wire is secure in the slot.
- Wire the Connection cable to the Forma 4100 Series Incubator:
- Power off the Thermo incubator. (CAUTION: Failure to do so may result in electrical damage or fire.)
- Open the cabinet door on the front of the incubator and remove the four screws as shown in Figure 2. Figure 2: Cabinet door on front of the incubator
- Remove the cover panel. Carefully slide out the electronics drawer and locate the Analog Output board.
- Route the Connection cable through the Analog Wiring Inlet to the Forma 4100 Series Incubator as shown in Figure 3. Figure 3: Routing for Connection cable through Analog Wiring Inlet
- Slightly loosen the screws for the TEMP, CO2, O2, and RH channels on the terminal block of the Thermo Forma 4100 Series Incubator, as shown in Figure 4.Figure 4: Connection Cable is wired into the TEMP, CO2, and RH (+ ) and (- ) Channels of the Thermo Incubator
- Insert the BLUE Wire from Connection Cable into the RH (-) slot. Tighten the screw to ensure that the wire is secure.
- Insert the ORANGE Wire from Connection Cable into the RH (+) slot. Tighten the screw to ensure that the wire is secure.
- Insert the YELLOW Wire from Connection Cable into the O2 (-) slot. Tighten the screw to ensure that the wire is secure.
- Insert the GREEN Wire from Connection Cable around the O2 (+) slot. Tighten the screw to ensure that the wire is secure.
- Insert the BROWN Wire from Connection Cable around the CO2 (-) slot. Tighten the screw to ensure that the wire is secure.
- Insert the BLACK Wire from Connection Cable around the CO2 (+) slot. Tighten the screw to ensure that the wire is secure.
- Insert the WHITE Wire from Connection Cable around the TEMP (-) slot. Tighten the screw to ensure that the wire is secure.
- Insert the RED Wire from Connection Cable around the TEMP (+) slot. Tighten the screw to ensure that the wire is secure.
- Re-install the cover panel. Replace the four screws on the cabinet door.
- Power on the Element-C.
- Power on the Thermo incubator.
- Confirm that incubator conditions appear on the Elemental Machines dashboard. Setup any desired alerts on the dashboard.